BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for Ultraview SL 2200, 2400, 2600, 2700 and XPREZZON Monitors with Perioperativ-Option, Spacelabs Medical

Product group Medical electronics / Electromedical devices
Reference 03656/11

Spacelabs Medical has issued a Field Safety Notice about a potential hazard associated with the use of patient monitors models 91367, 91369, 91370, 91387 and 91393.
The alarm tones are off if the device has been left in the END CASE state.