BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Field Safety Notice concerning a corrective action of the electrocardiogram (ECG) Screening software, Spacelabs Healthcare

Product group Medical electronics / Electromedical devices
Reference 05363/11

Spacelabs Healthcare has issued a Field Safety Notice concerning a corrective action of the electrocardiogram (ECG) Screening software version 3.31. If Lifescreen is left open and a new recording is loaded from Sentinel Cardiology Information Management System the file can get corrupted or a wrong ECG can occur. This can only happen if both recordings are greater than 32 hours long.

Field Safety Notice concerning a corrective action of the electrocardiogram (ECG) Screening software, Spacelabs Healthcare , Download_VeroeffentlichtAm_EN PDF, 115KB, File does not meet accessibility standards