BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for the STERRAD® 200 System, ASP

Product group Medical electronics / Electromedical devices
Reference 00477/12

Advanced Sterilization Products (ASP) has issued a field safety notice concerning with vacuum pump in the STERRAD® 200 System. It was determined that the vacuum pump may cause the emission of a slightly increased level of hydrogen peroxide (H2O2) vapour into the surrounding environment if not regularly serviced or replaced.

Safety Notice for the STERRAD® 200 System, ASP , Download_VeroeffentlichtAm_EN PDF, 46KB, File does not meet accessibility standards