Safety Information for monopolar electrodes mentioned in table 1, Reger Medizintechnik GmbH
2012.10.26
Product group Medical electronics / Electromedical devices - electrotherapy
Reference 01016/12
The company Reger Medizintechnik GmbH notified that the shafts of the electrodes mentioned in table 1 can heat up, when they are used in combination with radio-frequency generators (frequency higher than 1 MHz). The electrodes could affect a redness of skin, irritation or burns if the shaft contacts human tissue.