Safety Notice concerning a corrective action of the defibrillator/pacer SLIM unit, part of the Corpuls3 system, GS Elektromedizinische Geräte G. Stemple GmbH
2014.02.06
Product group Medical electronics / Electromedical devices - electrotherapy
Reference 00392/14
GS Elektromedizinische Geräte G. Stemple GmbH has issued a Field Safety Notice concerning a corrective action of the defibrillator/pacer SLIM unit, part of the Corpuls3 system. One component on a circuit board of units, bearing the serial numbers from 12800103 to 13851242 delivered between 07/2012 and 12/2013, could be damaged mechanically.
As a result, there is the risk that the radio connection between the modules is impaired.