Urgent Safety Notice for Patient Monitors B20 / B40 V1, B30 and B40 V2, GE Healthcare
2014.01.29
Product group Medical electronics / Electromedical devices - electrotherapy
Reference 07961/13
The B20, B30 and B40 Patient Monitors set the ECG filter setting incorrectly if the monitor starts with impedance respiration set in the ON position.