BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Lot Recall for "Monoaxial pedicular screws" Odalys, KISCO

Product group Non-active implants
Reference 01100/11

Monoaxial pedicular screws, Odalys: The company KISCO has informed about a lot recall of these products. The action was initiated because of a performed modification of the screwdriver tip (T form); concerned screws are no more compatible with the screwdriver.

Lot Recall for "Monoaxial pedicular screws" Odalys, KISCO , Download_VeroeffentlichtAm_EN PDF, 39KB, File does not meet accessibility standards