Lot Recall for the Targon PF Nail, AESCULAP
2012.08.31
Product group Non-active implants - bone surgery
Reference 05280/12
Targon PF Nail: The company AESCULAP has informed about a lot recall of these bone nails. The action was initiated because of a thread-error in such a way that it is not possible to torque the holding screw inside the nail.
Lot Recall for the Targon PF Nail, AESCULAP , Download_VeroeffentlichtAm_EN