BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Information concerning the Raphael ventilator from Hamilton

Product group Surgical equipment/ Anaesthesia
Reference 01654/06

Hamilton has issued an information concerning the Raphael ventilator. It’s alarm system can be put into a state where no visible or aqudiple patient alarms are triggered, if an oxygen cell calibration procedure is performed without compressed air supply.

Information concerning the Raphael ventilator from Hamilton , Download_VeroeffentlichtAm_EN PDF, 139KB, File does not meet accessibility standards