BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Information for the Centiva/5 and the Centiva/5 Plus critical care ventilators from GE Healthcare

Product group Surgical equipment/ Anaesthesia
Reference 03251/06

GE Healthcare has issued the information that the Centiva/5 and the Centiva/5 Plus critical care ventilators may shutdown if two independent anomalies occur. GE Heathcare will fix this problem by softwareupdate and installation of a new hardware circuit.

Information for the Centiva/5 and the Centiva/5 Plus critical care ventilators from GE Healthcare , Download_VeroeffentlichtAm_EN PDF, 182KB, File does not meet accessibility standards