BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Information concerning the Tracheostomy Tube Holders 240 Blue and 242 Blue from DALE Medical Products

Product group Surgical equipment/ Anaesthesia
Reference 03458/07

DALE Medical Products Inc. informs their customers per customer letter about risks of misuse of the 240 Blue and 242 Blue Tracheostomy Tube Holder. Altering the product fastener tabs by cutting or trimming is a contraindication not recommended. This alteration can weaken the stich integrity and cause the tab to break and detach and potentially lead to unintended decannulation.

Information concerning the Tracheostomy Tube Holders 240 Blue and 242 Blue from DALE Medical Products , Download_VeroeffentlichtAm_EN PDF, 193KB, File does not meet accessibility standards