BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for table brumaba L/Standard and model 2002, Brumaba

Product group Surgical equipment/ Anaesthesia
Reference 01039/08

Brumaba has issued a safety notice concerning table brumaba L/Standard and model 2002. The manufacturer became aware that high-frequency radiation from radio frequency surgery devices could let to unintentional movements of the table. A conversion of the foot switch is recommended. Advisory Notice is only available in German language.

Safety Notice for table brumaba L/Standard and model 2002, Brumaba , Download_VeroeffentlichtAm_EN PDF, 126KB, File does not meet accessibility standards