BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Recall RiwoMobile, Richard Wolf GmbH

Product group Surgical equipment/ Anaesthesia
Reference 01829/08

The Richard Wolf GmbH issued an information concerning a recall of the RiwoMobile equipment trolley. A production lot of the power module used in the affected trolley model 32114.001 is electrically falsely-dimensioned, that in case of a failure it would no longer be possible to switch on the power supply on the main switch; consequently all devices installed would remain functionless.

Recall RiwoMobile, Richard Wolf GmbH , Download_VeroeffentlichtAm_EN PDF, 124KB, File does not meet accessibility standards