Safety Notice for Acrobat 2000, Ondal Industrietechnik GmbH

Product group Surgical equipment/ Anaesthesia
Reference 02060/09

Ondal Industrietechnik GmbH has issued a Field Safety Notice concerning a corrective action of the spring-loaded arms Acrobat 2000 produced from 2000 to 2006. There is the risk that a welded seam at the pivot breaks and the light head falls down. The customers are requested to check the affected spring arms on the basis of the “Instructions for checking the front pivot on the Acrobat 2000 spring arm, Year of production 2000 – 2006” which are enclosed to the Safety Notice.