Safety Notice for Alaris Gateway Workstation, CareFusion

Product group Surgical equipment/ Anaesthesia
Reference 02408/09

CareFusion Corporation has issued a Field Safety Notice concerning a corrective action of the Alaris Gateway Workstation AGW. The clamp turn screw of the pole clamps with a T-bar handle with bull nosed thread end may be permanently bent (i.e. due to position change of the devices without loosening the clamp turn screw) and may result in allowing the devices to come away from the pole. The end users are recommended to take safety measures and they are provided with replacement pole clamps with an improved design. Advisory Notice is only available in German language.