BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Field Safety Notice concerning a corrective action of the TruPort ceiling support systems, Trumpf Medizin Systeme GmbH & Co. KG

Product group Surgical equipment/ Anaesthesia
Reference 03820/11

Trumpf Medizin Systeme GmbH & Co. KG has issued a Field Safety Notice concerning a corrective action of the TruPort ceiling support systems. An incorrectly installed electrical connection can cause higher transfer resistance, resulting in temperature increase and melting of plastic components. All potentially affected ceiling support systems will be checked and reworked, if necessary.

Field Safety Notice concerning a corrective action of the TruPort ceiling support systems, Trumpf Medizin Systeme GmbH & Co. KG , Download_VeroeffentlichtAm_EN PDF, 60KB, File does not meet accessibility standards