Safety Notice for the Acrobat 3000 spring arm, Trumpf Medizin Systeme GmbH + Co. KG

Product group Surgical equipment/ Anaesthesia - surgical equipment
Reference 03167/12

Trumpf Medizin Systeme GmbH + Co. KG has issued a Field Safety Notice concerning a corrective action of the Acrobat 3000 spring arm as support arm for iLED operating lights. Cracks can arise on the welded seam on the rear joint. This can result in support plates coming loose, and the spring arm suddenly falling down. The potentially faulty components in all affected spring arms will be exchanged.