Safety Notice for the Sorin S5/C5 Perfusion System, Sorin Group

Product group Surgical equipment/ Anaesthesia - surgical equipment
Reference 05818/12

Sorin Group has issued a Field Safety Notice concerning a problem of a “Fault in motor controller” alarm message. It was determind, that some of these events were the result of an electrical component that was out of specification.

Safety Notice for the Sorin S5/C5 Perfusion System, Sorin Group , Download_VeroeffentlichtAm_EN PDF, 559KB, File does not meet accessibility standards