BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Corrective action of the examination lights ML 301 which feature the spring arm of the type Acrobat Swing of Ondal Medical Systems GmbH, Gebrüder Martin GmbH & Co. KG

Product group Surgical equipment/ Anaesthesia - surgical equipment
Reference 00434/13

Gebrüder Martin GmbH & Co. KG has issued a Field Safety Notice concerning a corrective action of the examination lights ML 301 which feature the spring arm of the type Acrobat Swing of Ondal Medical Systems GmbH. There is the risk that the plastic joint breaks and the spring arm falls suddenly and without hindrance. Spring arms with plastic joints must not be used anymore and have to be replaced.

Corrective action of the examination lights ML 301 which feature the spring arm of the type Acrobat Swing of Ondal Medical Systems GmbH, Gebrüder Martin GmbH & Co. KG , Download_VeroeffentlichtAm_EN PDF, 6MB, File does not meet accessibility standards