BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Corrective action of the KLS Martin Double-TFT monitor support arm system, Gebrüder Martin GmbH & Co. KG

Product group Surgical equipment/ Anaesthesia - surgical equipment
Reference 01188/13

Gebrüder Martin GmbH & Co. KG has issued a Field Safety Notice concerning a corrective action of the KLS Martin Double-TFT monitor support arm system. The two M6 retaining screws between the crossbeam and receptacle pin can dissolve, with the result that the crossbeam and the monitors sink down by around 10 cm. On the screws of the affected products a modification has to be performed.

Corrective action of the KLS Martin Double-TFT monitor support arm system, Gebrüder Martin GmbH & Co. KG , Download_VeroeffentlichtAm_EN PDF, 4MB, File does not meet accessibility standards