Safety Notice concerning a corrective action of the spring arm of the type Acrobat Swing with a plastic joint and a carrying capacity of 2-3,3 kg, Ondal Medical Sytems GmbH

Product group Surgical equipment/ Anaesthesia - surgical equipment
Reference 05470/12

Ondal Medical Sytems GmbH has issued a Field Safety Notice concerning a corrective action of the spring arm of the type Acrobat Swing with a plastic joint and a carrying capacity of 2-3,3 kg. There is the risk that the joint breaks and the spring arm falls down abruptly and without interruption. It is recommended strictly avoiding the presence of any persons below the concerned devices and replacing the spring arms immediately.

Safety Notice concerning a corrective action of the spring arm of the type Acrobat Swing with a plastic joint and a carrying capacity of 2-3,3 kg, Ondal Medical Sytems GmbH , Download_VeroeffentlichtAm_EN PDF, 60KB, File does not meet accessibility standards