BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Corrective action of the S5 Perfusion System, Sorin

Product group Surgical equipment/ Anaesthesia - surgical equipment
Reference 05641/13

Sorin Group Deutschland GmbH has issued a Field Safety Notice concerning a corrective action of the S5 Perfusion System. An anomaly in firmware versions R3.13, R3.14 and R3.15 may result in failing of the automatic cardioplegia delivery function. An updated validated firmware corrects the problem.

Corrective action of the S5 Perfusion System, Sorin , Download_VeroeffentlichtAm_EN PDF, 904KB, File does not meet accessibility standards