BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for the Equipment Carts and Isolation Transformers on behalf of the company Trumpf Medizinsysteme, Karl Storz GmbH & Co. KG

Product group Surgical equipment/ Anaesthesia - surgical equipment
Reference 00857/14

On behalf of the company Trumpf Medizinsysteme GmbH & Co. KG the Karl Storz GmbH & Co. KG has issued a Field Safety Notice concerning a corrective action of the Equipment Carts 29005DRB, -DRS und -GI as well as the Isolation Transformers 29005TBG and 29003TBK. Due to damaging in course of the manufacturing process the protective conductor may have lost contact at the IEC receptacle for devices. In case of an insulation defect in an electrical device there is thus the risk of an electric shock. The main cords have to be exchanged.

Safety Notice for the Equipment Carts and Isolation Transformers on behalf of the company Trumpf Medizinsysteme, Karl Storz GmbH & Co. KG , Download_VeroeffentlichtAm_EN PDF, 1MB, File does not meet accessibility standards