Safety Notice concerning a corrective action of the Dräger monitor holders of the operating light systems Sola 500/700, Dräger

Product group Surgical equipment/ Anaesthesia - surgical equipment
Reference 01833/14

Dräger Medical GmbH has issued a Field Safety Notice concerning a corrective action of the Dräger monitor holders of the operating light systems Sola 500/700. Regarding ceiling-mounted installations with Ondal AC 2000 spring loaded arms shipped up until December 2006, is the risk that a welded seam breaks and the monitor falls down. The affected spring loaded arms will be replaced.