BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Information concerning an update of the WaveScan WaveFront software to version 3.671 from Advanced Medical Optics, Inc.

Product group ophthalmological technology
Reference 02179/07

The company Advanced Medical Optics, Inc. (AMO) issued an information concerning an update of the WaveScan WaveFront software to version 3.671. Two errors in the system software may result in an erroneous LASIK treatment calculation for patients.

Information concerning an update of the WaveScan WaveFront software to version 3.671 from Advanced Medical Optics, Inc. , Download_VeroeffentlichtAm_EN PDF, 75KB, File does not meet accessibility standards