Recall for Zyoptix® XP Microkeratome Motors, Bausch & Lomb

Product group ophthalmological technology
Reference 01733/08

The company Bausch & Lomb Inc. conducts a recall of a defined number of Zyoptix® XP Microkeratome Motors. There is the potential for the plastic tips on the drive shaft of the motors to come off the shaft and cause the blade to stop oscillating.