Software correction for Perimeter Octopus 900 with Flash Software Version V 2.1.4 and V 2.1.5, Haag-Streit AG
2011.11.16
Product group ophthalmological technology
Reference 05049/11
The company Haag-Streit AG notified of a software correction required for Perimeter Octopus 900 with Flash Software Version V 2.1.4 and V 2.1.5. Due to a software error the background illumination of the cupola may not be switched on at the start of an examination leading to results “better” than the actual ones.