B.Braun/Aesculap recalls Endoscopy Trolleys
2006.03.31
Product group Optics / Precision engineering
Reference 02252/05
The company B.Braun/Aesculap AG notiefied a recall of the Endoscopy Trolley type PV798, PV799, PV998 and PV999 due to a possible risk of breaking wheels. The trolleys will be equipped with a new version of wheels prepared with special glue that avoid a loosing of the threaded wheel bolts.