BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for DYONICS BONECUTTER ELECTROBLADE, Smith & Nephew

Product group Optics / Precision engineering
Reference 02924/10

Smith & Nephew has issued a Field Safety Notice concerning amended instructions for use for DYONICS BONECUTTER ELECTROBLADE (Cat. No. 7205962 and 72202213). If irrigant flow is insufficient, heating of the irrigant within the joint may occur, which may cause patient burns as the irrigiant exists one of the arthroscopic portals.

Safety Notice for DYONICS BONECUTTER ELECTROBLADE, Smith & Nephew , Download_VeroeffentlichtAm_EN PDF, 708KB, File does not meet accessibility standards