BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Information about UV7001 Therapy systems of H. Waldmann

Product group Physical therapy
Reference 02711/05

H. Waldmann GmbH & Co. KG informs user of UV7001 Theraphy Systems without service contract about the importance of the annual preventive maintenance. A white layer or staining on the surface of filter glasses reduces the permeability of UV-light and can lead to measuring errors and overdosing of the patient.

The advisory notice is only in German language available.

Information about UV7001 Therapy systems of H. Waldmann , Download_VeroeffentlichtAm_EN