Safety Notice concerning a corrective action of the infant incubator V-2200, Atom Medical

Product group Physical therapy
Reference 02484/10

Atom Medical has issued a Field Safety Notice concerning a corrective action of the infant incubator V-2200. If power supply should be interrupted the operation of neither the oxygen controller nor the humidity control function will be resumed automatically when the power supply has returned. The users are advised that the appropriate switches have to be pressed again in order to reactivate these control functions.