BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Recall of power supplies for Akita Jet and Akita II Apixneb inhalation devices, Activaero GmbH

Product group Physical therapy - inhalation therapy
Reference 02500/12

Activaero GmbH has issued a Field Safety Notice concerning a recall of power supplies manufactured by Friwo and delivered together with the Activaero Akita Jet and Akita II Apixneb inhalation devices. The affected power supplies can overheat under normal operating conditions, resulting in melted, charred or damaged plastic housing. There is the risk of fire, burns or electric shock.

Recall of power supplies for Akita Jet and Akita II Apixneb inhalation devices, Activaero GmbH , Download_VeroeffentlichtAm_EN PDF, 81KB, File does not meet accessibility standards