BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice concerning a corrective action of the incubators Giraffe and Giraffe Omnibed, GE Healthcare

Product group Physical therapy - climatotherapy
Reference 00323/13

GE Healthcare has issued a Field Safety Notice concerning a corrective action of the incubators Giraffe and Giraffe Omnibed. Devices with Servo Oxygen Control System manufactured from August 1, 2012 through December 20, 2012 may randomly overwrite the oxygen set points or radiant heater set points when powering on or recovering from a power failure event. The display printed circuit boards of the affected units will be exchanged.

Safety Notice concerning a corrective action of the incubators Giraffe and Giraffe Omnibed, GE Healthcare , Download_VeroeffentlichtAm_EN PDF, 66KB, File does not meet accessibility standards