Information for Brilliance BigBore CT scanner, Philips
2007.11.12
Product group Radiological technology
Reference 03911/07
Philips Medizin Systeme GmbH issued an information concerning Brilliance BigBore CT scanner with Tumor Localization application (Tumor LOC) using software V 2.2.1 through V2.2.5. Maximum Intensity Projection (MIP) images produced from respiratory gated images by Tumor Localization (Tumor LOC) will have incorrect Left/Right labels for certain patient orientations and view conventions.