BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Information for Omega IV and Omega V tables, GE Healthcare

Product group Radiological technology
Reference 00840/08

GE Healthcare has issued a customer letter about the risk of sudden table drop on Omega IV and Omega V tables on Advantx-E as well as on Innova 2000, 2100IQ, 3100/3100IQ und 4100/4100IQ cardiovascular X-ray systems. Only tables, which were manufactured between July 2003 and January 2007, are affected by this issue.

Information for Omega IV and Omega V tables, GE Healthcare , Download_VeroeffentlichtAm_EN PDF, 80KB, File does not meet accessibility standards