Reference 02499/08

Sirona Dental Systems GmbH has issued a Safety Notice concerning GALILEOS 3D X-ray system with GALAXIS software ≤ 1.5.x. If a CEPH ap or CEPH pa view is generated and printed via GALAXIS or SIDEXIS, the printout shows an incorrect side marker. A software update is planned.