BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for OEC Fluorostar 7900, GE Healthcare

Product group Radiological technology
Reference 01312/09

GE Healthcare has issued a Field Safety Letter concerning a radiation safety issue on X-ray system OEC Fluorostar 7900 with software version below 02.05.10. There is no audible alarm after 5 minutes of accumulated radiation time. A software update is available.

Safety Notice for OEC Fluorostar 7900, GE Healthcare , Download_VeroeffentlichtAm_EN PDF, 66KB, File does not meet accessibility standards