BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for NX software with CMS option, Agfa HealthCare

Product group Radiological technology
Reference 02275/09

Agfa HealthCare has issued a Field Safety Notice concerning a problem with “NX“ -Modality Workstation when used in combination with the optional “CMS“ Central Monitoring System. A user action on CMS may cause NX to unwantedly switch to another selected patient study.

Safety Notice for NX software with CMS option, Agfa HealthCare , Download_VeroeffentlichtAm_EN PDF, 118KB, File does not meet accessibility standards