Information for the Digital Image Capture System SDC HD 0240050888 and 0240050888i, Stryker
2009.12.21
Product group Radiological technology
Reference 04714/09
Stryker Endoscopy has issued a Field Safety Notice concerning a corrective action of the Digital Image Capture System SDC HD 0240050888 and 0240050888i. After 31th December 2009 saved files will not be assigned with the correct time/date and the system date will re-start from 1st October 2001. The affected devices should not be used until a software upgrade will be carried out.