BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Field Safety Notice concerning Hologic StereoGuide Biopsy System, Medicor

Product group Radiological technology
Reference 02936/10

Medicor has issued a Field Safety Notice concerning Hologic StereoGuide Biopsy System (also identified as ABBI and MultiCare) because of the potential of inadvertent stage movement if fluid enters into the System Keypad Control Panel. Users are requested to cover the control panel with a protective plastic cover.

Field Safety Notice concerning Hologic StereoGuide Biopsy System, Medicor , Download_VeroeffentlichtAm_EN PDF, 914KB, File does not meet accessibility standards