BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Recall of double foot switch KF21, Ziehm

Product group Radiological technology
Reference 00533/11

Ziehm has issued a Field Safety Notice concerning the recall of double foot switch KF21 (Ziehm material number 20125) manufactured between April and October 2010. There is a risk of unintentional radiation exposure because the pedal of the foot switch may not become inactive when the pedal is released. The following C-arms are affected: 8000, Compact, Solo, Vision, Vision FD, Vision FD Vario 3D, Vision RFD, Vision R, Vision Vario 3D und Vista plus.

Recall of double foot switch KF21, Ziehm , Download_VeroeffentlichtAm_EN PDF, 534KB, File does not meet accessibility standards