BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for Senographe DS and Senographe Essential Mammography systems, GE Healthcare

Product group Radiological technology
Reference 03353/11

GE Healthcare has issued a Field Safety Notice concerning potential inaccurate measurements on GE magnified images, acquired on Senographe DS and Senographe Essential Mammography systems with Magstand, when reviewed on workstations or PACS that are not adapted to GE Imager Pixel Spacing coding. A software update is planned.

Safety Notice for Senographe DS and Senographe Essential Mammography systems, GE Healthcare , Download_VeroeffentlichtAm_EN PDF, 87KB, File does not meet accessibility standards