BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for the Discovery NM/CT 670 SPECT/CT and Discovery NM 630 SPECT Systems, GE Healthcare

Product group Radiological technology - equipment for nuclear medical functional and localisation diagnostics
Reference 06607/12

GE Healthcare has issued a Field Safety Notice concerning a potential collision between table pallet and detector on Discovery NM/CT 670 SPECT/CT and Discovery NM 630 SPECT Systems. It is recommended to use the handheld controller at scan set-up. A software update is planned.

Safety Notice for the Discovery NM/CT 670 SPECT/CT and Discovery NM 630 SPECT Systems, GE Healthcare , Download_VeroeffentlichtAm_EN PDF, 30KB, File does not meet accessibility standards