BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Software Upgrade for DigitalDiagnost X-ray systems with 3.0.x Software, Philips Medical Systems

Product group Radiological technology - stationary radiological equipment
Reference 07169/12

Philips Medical Systems has issued a Field Safety Notice concerning a software upgrade required on DigitalDiagnost X-ray systems with software versions 3.0.1, 3.0.2, 3.0.3. Software release 3.0.3 contains a defect that may result in the incorrect placement of the electronic left/right marker indicating the patient side if an image is rotated.

Software Upgrade for DigitalDiagnost X-ray systems with 3.0.x Software, Philips Medical Systems , Download_VeroeffentlichtAm_EN PDF, 115KB, File does not meet accessibility standards