BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Software Upgrade required on DigitalDiagnost X-ray systems with software versions 2.0.2 und 2.1.3, Philips Medical Systems

Product group Radiological technology - stationary radiological equipment
Reference 07281/12

Philips Medical Systems has issued a Field Safety Notice concerning a software upgrade required on DigitalDiagnost X-ray systems with software versions 2.0.2 und 2.1.3. Software release 3.0.3 has been installed to modify the appearance of the mirror icon in release 2.0.2, but contains a defect that may result in the incorrect placement of the electronic left/right marker. The new software release 2.1.4 addresses both the original and new issue.

Software Upgrade required on DigitalDiagnost X-ray systems with software versions 2.0.2 und 2.1.3, Philips Medical Systems , Download_VeroeffentlichtAm_EN PDF, 103KB, File does not meet accessibility standards