Corrective action of the intraoral digital imaging plate system DIGORA Optime Type DXR-60, Soredex PaloDEx

Product group Radiological technology - auxiliary equipment and supplies for radiology / nuclear medicine
Reference 04617/13

Soredex PaloDEx has issued a Field Safety Notice concerning a corrective action of the intraoral digital imaging plate system DIGORA Optime Type DXR-60. In situations where the affected devices, serial numbers from SL1300505 to SL1300927, display an “error 3” message requiring service, there is a risk that the laser light inside the units illuminates continuously and the service technician is unexpectedly exposed to the laser beam.