BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Corrective action of the Treatment Planning Systems Eclipse from Varian

Product group Radiotherapy / Radiation protection
Reference 00274/05

Varian Medical Systems informed about a corrective action of the Treatment Planning Systems Eclipse using Pencil Beam Convolution (photons) and Generalized Gaussian Pencil Beam (electrons) dose calculation algorithms build 7.2.24 or earlier. Field dose distributions may be rotated at certain couch and gantry angle combinations.

Corrective action of the Treatment Planning Systems Eclipse from Varian , Download_VeroeffentlichtAm_EN PDF, 94KB,