BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Anomaly in Eclipse Photon Single Pencil Beam Convolution from Varian

Product group Radiotherapy / Radiation protection
Reference 02653/05

Varian informs about an anomaly in Eclipse Photon Single Pencil Beam Convolution (PBC) version 7.2.35. Truncated dose distributions for tangential fields may occur if the X1 or Y1 jaws are in an over-travel position.

Anomaly in Eclipse Photon Single Pencil Beam Convolution from Varian , Download_VeroeffentlichtAm_EN PDF, 157KB,