Information concerning the interface mount by Clinacs with Type 3 accessories from Varian

Product group Radiotherapy / Radiation protection
Reference 02278/07

Varian medical systems has issued an information concerning installation failure of the interface mount by Clinacs with Type 3 accessories. The wedge tray may slip out of the interface mount when the gantry is rotated in a position where the opening of the interface mount is pointed in the direction of the floor.

Information concerning the interface mount by Clinacs with Type 3 accessories from Varian , Download_VeroeffentlichtAm_EN PDF, 104KB, File does not meet accessibility standards