Information for Eclipse Client 8.1.18 and DCF 8.1.17, Varian
2008.04.01
Product group Radiotherapy / Radiation protection
Reference 00742/08
Varian Medical Systems has issued an information concerning a problem with the calculation of IRREG plans containing wedges with Eclipse Client 8.1.18 and DCF 8.1.17. An anomaly was detected such that, for a wedged field of an IRREG plan, the calculation will ignore the wedge and calculate the field as open.
Information for Eclipse Client 8.1.18 and DCF 8.1.17, Varian , Download_VeroeffentlichtAm_EN